Beer-Lambert Law Calculator

The Beer-Lambert law links light absorbance to solution concentration through A = ε·c·l. Choose what to solve for, enter the other three quantities, and this calculator returns the result along with transmittance. Unit switches for concentration (mol/L, mmol/L, umol/L) and path length (cm, mm, m) are built in, and a transmittance converter mode handles %T input directly.

Your details

Beer-Lambert solves for any of the four variables. Transmittance converter converts between %T, T, and A.
Choose the unknown. Supply the remaining three quantities below.
Dimensionless absorbance from the spectrophotometer. Ignored when solving for absorbance.
Also called the molar extinction coefficient. Specific to the analyte and wavelength.
L·mol⁻¹·cm⁻¹
Molar concentration of the absorbing species. Switch units to mmol/L or umol/L for dilute samples.
Distance light travels through the sample. Standard cuvettes are 1 cm. Switch to mm for microvolume cuvettes.
Result
c = 5.0000e-5 mol/L

The quantity you chose to solve for.

Absorbance (A)0.5
Transmittance (%T)31.62%
Transmittance (T)0.3162
Concentration (c)0.00005mol/L
0.5 A
Very low<0.1Linear range0.1-1Marginal1-2Too high2+
00.51050100
Concentration (umol/L)
Absorbance (A)
Concentration (umol/L)Calibration curve (A = ε·l·c)
00
100.1
200.2
300.3
400.4
500.5
600.6
700.7
800.8
900.9
1001

c = 5.0000e-5 mol/L, with 31.62% of light transmitted.

  • An absorbance of 0.5 is in the optimal linear range for Beer-Lambert accuracy.
  • 31.62% of the incident light passes through - the remaining 68.38% is absorbed.
  • Absorbance is directly proportional to both concentration and path length - doubling either doubles A.
  • For most reliable results keep absorbance between 0.1 and 1.0, diluting the sample if needed.

Next stepBuild a calibration curve of absorbance vs. known concentrations to read off unknowns from the slope (ε·l).

Formula

A=ε⋅c⋅l,%T=100×10−A,A=2−log⁡10(%T)A = \varepsilon \cdot c \cdot l, \qquad \%T = 100 \times 10^{-A}, \qquad A = 2 - \log_{10}(\%T)

Worked example

With ε = 10,000 L·mol⁻¹·cm⁻¹, measured absorbance A = 0.5, and a 1 cm cuvette: c = A / (ε·l) = 0.5 / (10,000 × 1) = 5 × 10⁻⁵ mol/L (50 umol/L). Transmittance: 100 × 10⁻⁰·⁵ = 31.6%. If the spectrophotometer reports %T = 25%, then A = 2 - log10(25) = 0.602.

What the Beer-Lambert law describes

The Beer-Lambert law (also called Beer's law or the Beer-Lambert-Bouguer law) states that the absorbance of light by a solution is directly proportional to the concentration of the absorbing species and to the distance the light travels through the sample. Written as A = ε·c·l, absorbance A is a dimensionless number, ε is the molar absorptivity (molar extinction coefficient) in L·mol⁻¹·cm⁻¹, c is the molar concentration in mol/L, and l is the optical path length in centimetres. Because the three factors multiply, knowing any three lets you solve for the fourth. This is the foundation of UV-Vis spectrophotometry: a single light-intensity measurement becomes a quantitative concentration.

Absorbance, transmittance, and the converter

A spectrophotometer measures transmittance, the fraction of incident light that exits the sample, and converts it internally to absorbance via A = -log10(T). In percent transmittance, that becomes A = 2 - log10(%T). An absorbance of 0 means all light passes (100% T); A = 1 means only 10% passes; A = 2 means 1% passes. Absorbance is preferred for analytical work because it is linear in concentration, making calibration curves straight lines. The transmittance converter mode on this calculator handles any of the three forms: you can enter %T, fractional T, or A and get the others instantly.

Unit switches: mmol/L, umol/L, mm path length

Real-world samples span a wide range of concentrations. Protein assays often work in the micromolar range (umol/L), while simple dye solutions may sit in the millimolar (mmol/L) range. This calculator lets you enter concentration in mol/L, mmol/L, or umol/L and converts internally before solving. Similarly, microvolume cuvettes and flow cells use path lengths of 1 mm or shorter, so path length can be entered in cm, mm, or m. The molar absorptivity ε always stays in L·mol⁻¹·cm⁻¹ (the SI analytical chemistry convention) and the conversions keep everything consistent.

When the linear law breaks down

Beer's law assumes monochromatic light, a homogeneous and dilute solution, and absorbers that act independently. In practice the linear relationship holds well only at low absorbances, typically below about 1.0, and bends at higher concentrations because molecules begin to interact, the refractive index shifts, and stray light reaches the detector. Chemical effects such as association, dissociation, or solvent reaction can also shift ε. For accurate work keep samples in the A = 0.1 to 1.0 window, confirm linearity with a calibration curve, and if the reading exceeds 1, dilute by a known factor and multiply the result back.

Building a calibration curve

A single ε value from the literature gives a rough answer, but a real calibration curve is more reliable because it accounts for instrument-specific factors, solvent effects, and any deviation from pure Beer-Lambert behaviour. Prepare a series of standards at known concentrations spanning your expected range, measure each absorbance, and plot A on the y-axis against concentration on the x-axis. The slope of the best-fit line equals ε·l. Divide any unknown sample's absorbance by that slope to get its concentration. The chart built into this calculator shows the expected straight-line relationship from your current ε and l values so you can visually check whether your measured point falls on the line.

Typical molar absorptivity values for common analytes

AnalyteWavelength (nm)ε (L·mol⁻¹·cm⁻¹)Notes
BSA protein (Bradford assay, dye-bound)595~70,000Brilliant Blue G dye complex
NADH / NADPH3406,220Standard for enzymatic assays
DNA (double-stranded)260~6,600 (per base pair avg.)Used for concentration and purity
Caffeine2729,800In aqueous solution
Potassium permanganate (KMnO4)5252,400Classic teaching example
p-Nitrophenol (ionised, pH > 9)40018,600Enzyme assay product
Hemoglobin (oxyHb)415125,000Soret band, per heme

ε values are approximate and wavelength-specific. Always use a calibration curve or a literature value at your exact wavelength.

Frequently asked questions

What units does the Beer-Lambert law use?

In the standard form A = ε·c·l, absorbance A is dimensionless, molar absorptivity ε is in L·mol⁻¹·cm⁻¹, concentration c is in mol/L, and path length l is in cm. With these units the products cancel to leave a unitless number. This calculator also accepts mmol/L or umol/L for c, and mm or m for l, converting to mol/L and cm internally before solving.

How do I find concentration from absorbance?

Rearrange A = ε·c·l to c = A / (ε·l). Read the absorbance from your spectrophotometer at the analyte's peak wavelength, look up the molar absorptivity for that compound, and divide by ε times the cuvette path length. For example, A = 0.5 with ε = 10,000 L·mol⁻¹·cm⁻¹ in a 1 cm cell gives c = 0.5 / 10,000 = 5 × 10⁻⁵ mol/L. In practice a calibration curve of known standards gives a more reliable result than a single literature ε.

Why should absorbance stay below about 1?

At A = 1, only 10% of the light reaches the detector; at A = 2, only 1% does. As transmittance shrinks, instrument noise and stray light make up a larger fraction of the tiny signal, so readings lose precision and the linear relationship bends. Keeping A between 0.1 and 1.0, diluting if necessary, gives the most accurate linear measurements. If a reading exceeds 1, dilute by a known factor and multiply the concentration result back up.

How do I convert between percent transmittance and absorbance?

Use A = 2 - log10(%T). For example, 25% T gives A = 2 - log10(25) = 2 - 1.398 = 0.602. Going the other way: %T = 100 × 10^(-A), so A = 0.602 gives %T = 100 × 10^(-0.602) = 25%. The Transmittance Converter mode on this calculator handles both directions.

What is molar absorptivity and where do I find it?

Molar absorptivity (ε), also called the molar extinction coefficient, is a measure of how strongly a compound absorbs light at a particular wavelength. It is unique to the molecule and the wavelength and is tabulated in the chemical literature, reagent kit documentation, and spectrophotometer software. Common reference values include ε = 6,220 L·mol⁻¹·cm⁻¹ for NADH at 340 nm, and ε = 125,000 L·mol⁻¹·cm⁻¹ for oxyhaemoglobin at 415 nm. If no literature value is available, measure a series of standards and calculate ε from the slope of the calibration curve: slope = ε·l.

Can I use this calculator for DNA or protein concentration?

Yes. For double-stranded DNA, a widely used approximation is that an absorbance of 1.0 at 260 nm in a 1 cm cuvette corresponds to about 50 ug/mL. For single-stranded DNA it is roughly 33 ug/mL, and for RNA about 40 ug/mL. These are concentration-by-mass conventions rather than strict Beer-Lambert with a single ε, but Beer's law still underpins the measurement. The A260/A280 ratio is also used to assess purity. For proteins measured directly at 280 nm, ε depends on the specific sequence (tryptophan and tyrosine content).

Sources

Written by Grace Mbeki, MSc Data Scientist & Educator · Nairobi, Kenya